Pressure Systems 9022 Bedienungsanleitung Seite 86

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 194
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 85
www.meas-
spec.com
Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
3 - 53
Example:
Send TCP/IP command to 9016 module (via its connected socket) that returns
decimal voltage data (of the raw temperature signal) for channels 1, 5, 9, and 13:
n11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
0.53013 0.541698 0.503633 0.000000
In this example channels 13, 9, and 5 return normal temperature voltage signals in
the range of 0.5 to 0.6 volts. Note that channel 1 returns a value of 0.0 volts,
indicating a possible error in its temperature signal.
A Model 9021/9022 example would be similar, but without specifying the
non-existent channel 13. (“n01110”)
Seitenansicht 85
1 2 ... 81 82 83 84 85 86 87 88 89 90 91 ... 193 194

Kommentare zu diesen Handbüchern

Keine Kommentare