Measurement Specialties, Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
5 - 4
In the process of performing general maintenance on a module and in printed-circuit board
replacement, the following tools may be required:
3/32" and 5/64" Allen-head screwdrivers,
a 3/16" hex wrench,
a needle nose tweezers,
a Phillips-head screwdriver,
a small adjustable wrench,
a tube of Krytox
®
lubricant (P/N 42-06-KRYX),
a spray bottle of silicone liquid lubricant (P/N
41-06-Silicone).
When performing any type of maintenance of NetScanner
System components, the following
guidelines and precautions should always be followed:
Verify that the work area and technicians are properly grounded to prevent
damage to electronic components due to electrostatic discharge.
Ensure that all electrical and pneumatic connections have been removed
from the module.
Ensure that the work area is free of dust and other possible contaminants that
may affect the high tolerance machined parts (and pneumatic seals, if model has
an integral manifold).
Care must be taken to prevent contaminants from reaching O-ring surfaces. If
O-ring surfaces require cleaning, use a lint-free applicator with acetone to
remove dirt and lightly lubricate the O-ring surface with Krytox® provided in
the maintenance kit.
Never use sharp objects to cut tubing from the bulged tubes. The tiny scratches
left on the tubes could cause leaks.
NOTE: MEAS offers an
optional maintenance kit (P/N
S901-0200000000) with the
lubricants included.