
www.meas-
Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
1 - 72
ii
dd
Description
‘q’ read
index
01
Retrieve user-provided parameters out of non-volatile
memory
27
00
Store current coefficient terms into transducer’s
non-volatile memory. (For use only with non-digitally
compensated 9400 sensors with EEPROM.) (‘w27’ selects
all channels) (‘w2700 eeee’ selects individual channels via
the position field)
3C
00
(9016 only) Set Temperature Range = 0 to 60ºC default
06
Set Temperature Range = -30 to 60ºC
07
Set Temperature Range = -20 to 70C
Modification of option 13 hex results in the new option selection becoming the module’s new
power-on default. All other options must be stored in non-volatile flash memory using ‘w07’
command in order to be retained after the module power cycles.
The Valve Position indexes (ii=12 and ii=0C) (for Model 9016 only) each have two states (00/01)
that when combined provide four (4) possible states of the C1/C2/C3/C4 internal valves. This
“Logical Rotary Switch” with four (4) positions is summarized in the following chart:
C1 Energized
C2 Not
(0C=01)
C2 Energized
C1 Not
(0C=00)
C3 Energized
C4 Not
(12=01)
PURGE
position
LEAK/CHECK
position
C4 Energized
C3 Not
(12=00)
CAL/RE-ZERO
position
RUN
position
Kommentare zu diesen Handbüchern