Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
3 - 17
CommandC’Sub-command Index 00: Configure & Start Multi-Point Calibration
This sub-command has four (4) additional required parameters used to configure and start the
Multi-Point Calibration function.
Command
C 00 pppp npts ord avg
‘C’ is the command letter.
‘ 00is the sub-command index (ii) for Configure & Start.
ppppis a 1-4 hex digit position field (channel selection bit map), that selects
any of the 1-16 (9016) or 1-12 (9021/9022) internal channels to be affected by
the multi-point calibration.
npts’ is the number of unique calibration points (between 1 and 19) to be
supplied during the calibration function.
ordis the order of the adjustment fit, which currently must be 1 for a 1st
order linear fit of the calibration data (i.e., a straight line).
avg is the number of A/D data samples collected and averaged for each
calibration point supplied (must be a power of 2 in the set 2, 4, 8, 16, or 32)
NOTE: all parameters are separated by a space.
Response
A
A’ is the acknowledge letter and is returned if all parameters are supplied with
reasonable values. Else, an error (‘N’) response is returned.
Description: Configures and starts the Multi-Point Calibration function. It specifies the
particular channels (pppp) whose offset and span adjustment coefficients will be
replaced when the function is completed. All specified channels must have the
same full-scale pressure range. Modules with more than one range of internal
transducers installed must have channels from each range calibrated separately.
This sub-command immediately alters the module’s normal data acquisition,
processes A/D samples for average count (default = 8, or as per the ‘w10dd
command), and uses the sub-command’s avg parameter sample count instead. A
larger count (e.g., 32) is encouraged for calibration purposes. The original sample
count will not be restored until the calibration function ends or is aborted (per
other sub-commands described on the following pages).
NOTE: Avoid confusing this Calibrate command ‘C’ (upper case C) with the
Configure/Control Autonomous Host Streams command ‘c’ (lower case c). Like “c,”
but unlike most other module commands, all sub-commands of this command require a
space between the command id (‘C’) and the first parameter (ii).