Pressure Systems 9022 Bedienungsanleitung Seite 59

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 194
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 58
Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
3 - 29
Example:
! Send TCP/IP command to 9016 module (via its connected socket) that returns
decimal raw “pressure” A/D counts data for channels 1, 5, 9, and 13:
a11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
32767.000000 -32700.000000 10.000000 16385.000000
Please note that channel 13 is saturated at +full scale and channel 9 is almost
saturated at -full scale. Channel 5 reads near zero and channel 1 is about ½ +full-
scale.
This command example also works for Models 9021 and 9022 if the non-existent
channel 13 is not set in the position field bit map.
Seitenansicht 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 193 194

Kommentare zu diesen Handbüchern

Keine Kommentare